Simple Locking Intramedullary (Slim) System Peg-28-036
Ref PEG-28-036Model PEG IMPLANT
Active
Valid until
Not listed
FDA recalls
1 ongoingDocuments
None linked in EUDAMED
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number7540194SLM-PEGN4Q | Date26 May 2026 | StatusActive |
| United States | GUDID listing GUDID | Number07540194011329 | Date19 Mar 2021 | StatusListed |
Certificates
EUDAMED lists no certificate for this device's Basic UDI-DI.
Recalls
-
- Ongoing Z-2941-2024
During the assembly of the FRT250 cartridge assembly, an incorrect component was used. Specifically, the FRT240-HEX (2) was used instead of the FRT250-HEX.
-
- Terminated Z-1546-2015
The hexagonal tip of the Female Driver instrument for the Fassier-Duval IM Telescopic System is 0.06 mm under the specifications. This may result in a reduction of the total strength of the instrument, possibly leading to deformation, wear, and tip breakage.
-
- Terminated Z-1472-2014
The outer package label contained the incorrect length of the device. There is a label on the outer package and an additional label on the product contained in the package.
-
- Terminated Z-1448-2014
The Male Components in this lot are made of material with lower strength than manufacturing specifications.
-
- Terminated Z-1494-2011
Possible breakage. Structural weakness has been detected on the Male Driver 3.2mm and 4.0mm sizes. In certain cases, this weakness could lead to a premature failure of the instrument if it is submitted to large torsional loads higher than the normal forces required to introduce the male component.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade Name
- The Simple Locking Intra Medullary System EUDAMEDSimple Locking Intramedullary (Slim) System GUDID
- Generic Name
- Orthopaedic bone screw (non-sliding)
- Model / Ref / Catalog
- PEG-28-036, PEG IMPLANT
- Primary DI / UDI-DI
- 07540194011329
- Basic UDI-DI
- 7540194SLM-PEGN4Q
- FDA Product Code
- HSB
- EMDN Code
- P09120601 Cortical screws
- GMDN Term
- Orthopaedic bone screw (non-sliding)
- Device Class
- Class IIb EUDAMEDClass II GUDID
- Regulation Number
- 888.3020
- Market of Registration
- Sweden
- GUDID distribution status
- In Commercial Distribution
- Sterilization Method
- Moist Heat or Steam Sterilization
- Size
- Outer Diameter — 2.8 MillimeterLength — 36 Millimeter
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device Class | Class IIb | Class II |
Other devices from Orthopediatrics Canada ULC
Properties
Other properties (20)
- Sterile: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No