Skip to content

SmartDrive® MX2+

by Max Mobility, LLC

Ref MX2+

SmartDrive MX2+ is a wheelchair electric-motor-driven propulsion system (EU MDR Class I; FDA Class II; FDA product code ITI, Wheelchair, Powered) made by Max Mobility, LLC, registered in EUDAMED on 8 Oct 2026 (Basic UDI-DI 08618960003SDMX237), cleared under FDA 510(k) K151199 on 30 Jul 2015 and listed in FDA GUDID since 16 Jan 2017 (UDI-DI 00861896000310), active as of 22 Oct 2026. 7 FDA recalls are linked to it.

Active European Union and United States. Also listed in GUDID (US).

CE MDR 510(k)

Valid until

No date given

CE: no date in EUDAMED. FDA: not time-limited

FDA recalls

7 ongoing

This device is named in the recall

Documents

None linked in EUDAMED

A group certificate may exist

510(k) K151199

FDA submission, from GUDID

Trust

Unverified

Source EUDAMED, GUDID, synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDR EUDAMED Number08618960003SDMX237 Basic UDI-DI 00861896000310UDI-DIDate8 Oct 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number00861896000310 UDI-DI Date16 Jan 2017ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE GUDID from GUDID Substantially Equivalent · Traditional NumberK151199 Date30 Jul 2015DecidedStatusActive

Official records

From the FDA record

Device description
The device consists of a motorized drive unit that interfaces with the wheelchair frame via removable mounting components. User input commands are provided through compatible control components, including a switch controller and/or Speed Control Dial (Part Numbers MX2 3DCK, MX2 3DC, MX2 3DCMC). The SpeedControl Dial is no longer commercially available. User input and device configuration may also be supported through compatible Smart Wearable applications operating on SmartDrive MX2+ WearOS and SmartDrive MX2+ WatchOS platforms, which interface with the SmartDrive system. The PushTracker mobile application, available on iOS and WearOS platforms, is used to configure device settings and select control input methods.
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 22 Oct 2026, as the labeller entered it.

Across registries

3 registries list this device

  • EUDAMED Basic UDI-DI 08618960003SDMX237 registered 8 Oct 2026 Active
  • FDA 510(k) K151199 decided 30 Jul 2015 Active
  • FDA GUDID UDI-DI 00861896000310 listed 16 Jan 2017 Listed

Recalls

7 recalls · 7 ongoing · FDA

  • 3 recalls, same reason

    • Ongoing Z-2538-2025 FDA Class I initiated 12 Aug 2025
    • Ongoing Z-2539-2025 FDA Class I initiated 12 Aug 2025
    • Ongoing Z-2540-2025 FDA Class I initiated 12 Aug 2025

    Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, their is a potential for lost of control of the device potentially resulting in minor or major injuries.

  • 2 recalls, same reason

    • Ongoing Z-1116-2025 FDA Class I initiated 20 Dec 2024
    • Ongoing Z-1117-2025 FDA Class I initiated 20 Dec 2024

    Due to faulty circuit board in speed control remote use in conjunction with a wheelchair power assist, their is a potential for lost of control of device potentially resulting in minor and major injuries.

  • 2 recalls, same reason

    • Ongoing Z-2004-2023 FDA Class II initiated 10 May 2023
    • Ongoing Z-2005-2023 FDA Class II initiated 10 May 2023

    When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.

Timeline

10 dated events

  1. Cleared FDA 510(k) K151199
  2. Listed FDA GUDID 00861896000310
  3. Recall initiated FDA recalls Z-2004-2023
  4. Recall initiated FDA recalls Z-2005-2023
  5. Recall initiated FDA recalls Z-1116-2025
  6. Recall initiated FDA recalls Z-1117-2025
  7. Recall initiated FDA recalls Z-2538-2025
  8. Recall initiated FDA recalls Z-2539-2025
  9. Recall initiated FDA recalls Z-2540-2025
  10. Registered EUDAMED 08618960003SDMX237

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDR Active
  • United States FDA 510(k) Active

Specification

EUDAMED · GUDID

Trade name
SmartDrive MX2+ EUDAMED
SmartDrive® GUDID
Generic name
Wheelchair electric-motor-driven propulsion system
Device name
MX2+
Model / reference
MX2+
Primary DI / UDI-DI
00861896000310
Basic UDI-DI
08618960003SDMX237
510(k) Number
FDA product code
ITI
EMDN code
Y122410 Electric propulsion unit
GMDN term
Wheelchair electric-motor-driven propulsion system
Device class
Class I EUDAMED
Class II GUDID
Regulation number
890.3860
EU registration type
Standard device
EU market of registration
Malta
GUDID distribution status
In Commercial Distribution

Truth per registry

2 sources · 1 field differs

FieldEUDAMEDGUDID
Device classEUDAMED
Class I
GUDID
Class II

Device class is stated on each registry's own scale: Class I in the EU (EUDAMED), Class II in the US (FDA GUDID). EU and US classes are different schemes, so this is not a disagreement.

Similar devices in this category

Other devices filed under ELECTRIC PROPULSION UNIT.

Other devices from Max Mobility, LLC

Properties

  • Over-the-counter
  • Serial number
  • Manufacturing date
Other properties (22, all answered No)