SnapshotGLO (KB100)
510(k) K242669
SnapshotGLO (KB100) is a Class II device made by Kent Imaging Inc, cleared under FDA 510(k) K242669 on 26 Mar 2025, active as of 11 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK242669 | Date26 Mar 2025 | StatusActive |
Official records
- FDA 510(k) database K242669
Recalls
-
- Ongoing Z-2775-2026
The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.
Timeline
- Cleared FDA 510(k) K242669
- Recall initiated FDA recalls Z-2775-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- QJF
- FDA applicant
- Kent Imaging, Inc.
- Device class
- Class II
- Regulation
- 21 CFR Part 878.4550
- Advisory committee
- General, Plastic Surgery
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K242669
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| SnapshotGLO KB100 | KB100AA / KB100 | B376KB1000 | — | On the market |