Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/Verio
by Siemens Medi Cal Solutions, Inc.
510(k) K121434
Active
510(k)
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK121434 | Date5 Nov 2012 | StatusActive |
Recalls
-
- Ongoing Z-2666-2025
- Ongoing Z-2665-2025
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
- Terminated Z-1308-2015
The gradient output supervision was permanently turned off on the MAGNETOM system, meaning that gradient outputs could exceed IEC60601-2-33 limits and peripheral nerve stimulation could occur.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LNH
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1000
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent