Skip to content

Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/Verio

by Siemens Medi Cal Solutions, Inc.

510(k) K121434

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK121434 Date5 Nov 2012ClearedStatusActive

Recalls

3 recalls · 2 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-2666-2025 FDA Class I initiated 28 Aug 2025
    • Ongoing Z-2665-2025 FDA Class I initiated 28 Aug 2025

    There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

    • Terminated Z-1308-2015 FDA Class II initiated 9 Mar 2015

    The gradient output supervision was permanently turned off on the MAGNETOM system, meaning that gradient outputs could exceed IEC60601-2-33 limits and peripheral nerve stimulation could occur.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LNH
Device Class
Class II
Regulation
21 CFR Part 892.1000
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent