SOMATOM Definition AS/AS+
by Siemens Medical Solutions USA, Inc.
510(k) K143400
SOMATOM Definition AS/AS+ is a medical device (FDA Class II; FDA product code JAK, System, X-Ray, Tomography, Computed) made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K143400 on 8 Apr 2015, active as of 22 Oct 2026. 5 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK143400 | Date8 Apr 2015 | StatusActive |
Official records
- FDA 510(k) database K143400
From the FDA record
- Received by the FDA
- 28 Nov 2014
- FDA decision
- 8 Apr 2015 (Substantially Equivalent)
- FDA review time
- 131 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Terminated Z-2440-2019
SOMATOM Definition Edge, SOMATOM Definition AS, SOMATOM Definition Flash with software syngo.CT with Software version VA48A_SP5 may result in scanning workflow interruptions and unexpected user notifications. and may result in a delay in diagnosis and/or patient rescans
-
- Terminated Z-0604-2017
A problem can occur with the small light marker windows, used for the positioning laser and are integrated in the front cover of the SOMATOM Definition AS and Definition Flash systems, by either becoming loose or potentially dropping out. There is a risk that a person could touch rotating or electrical parts of the gantry if they were to reach through the window opening.
-
- Terminated Z-1350-2016
Siemens became aware that the front cover on some Siemens CT system was not properly welded. It is possible for the welded-in threaded bolt to break out of the ground plate of the front cover if the cover is opened during service. This could not happen during clinical operation. Therefore, there is no potential risk for the user or patient and the CT System has no risk for a malfunction during clinical operation
-
- Terminated Z-1519-2016
Software bugs in these SOMATOM systems could possibly cause scan aborts, rescans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast medium, accidental x-radiation to the user or impede customer workflow.
-
- Terminated Z-0021-2016
software bug issues for SW-Version VA48A_SP0. The following safety issues were resolved: 1) Correction to improve visual warning and error indication son the gantry display. 2) Correction to improve acquisition data in order to optimize image quality. 3) Correction to improve robustness and general system behavior in some exception handling procedures. 4) Correction to improve auto post processin
Timeline
- Cleared FDA 510(k) K143400
- Recall initiated FDA recalls Z-0021-2016
- Recall initiated FDA recalls Z-1519-2016
- Recall initiated FDA recalls Z-1350-2016
- Recall initiated FDA recalls Z-0604-2017
- Recall initiated FDA recalls Z-2440-2019
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- JAK
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1750
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code JAK (System, X-Ray, Tomography, Computed).
- uCT 760 with uWS-CT-Dual Energy Analysis, uCT 780 with uWS-CT-Dual Energy Analysis by Shanghai United Imaging Healthcare Co., Ltd.
- MODIFICATION TO NEUVIZ DUAL MULTI-SLICE CT SCANNER SYSTEM by Neusoft Medical Systems Co., Ltd.
- FALCON DISK DRIVE SUBSYSTEM by Sokar, Inc.
- 3D ACCU-I-TOMO XYZ SLICE VIEW TOMOGRAPH, MODEL MCT-1 EXF by J. MORITA U.S.A., INC
- SYNGO VOLUME PERFUSION-CT NEURO by Siemens Medi Cal Solutions, Inc.
- Marie by Leo Cancer Care
- IMATRON ULTRA ACCESS WORKSTATION WITH CARDIAC SOFTWARE EXTENSIONS by Imatron, Inc.
- GYROVIEW-HR by Philips Medical Systems
- SHIMADZU WHOLE BODY X-RAY CT SCANNER SCT-3000TX by Shimadzu Precision Instruments, Inc.
- TECHNICARE 720 HPS CT SCANNER by Technicare Corp.