  "format": "devicerts.device.v1",
  "cite_as": "SOMATOM Definition AS/AS+, Siemens Medical Solutions USA, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/somatom-definition-as-as-jak (data as of 2026-11-18).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/somatom-definition-as-as-jak",
    "json_url": "https://devicerts.com/devices/somatom-definition-as-as-jak.json",
    "markdown_url": "https://devicerts.com/devices/somatom-definition-as-as-jak.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-18",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/somatom-definition-as-as-jak",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "SOMATOM Definition AS/AS+",
    "registry_name": "SOMATOM Definition AS/AS+",
    "url": "https://devicerts.com/devices/somatom-definition-as-as-jak",
    "manufacturer": {
      "name": "Siemens Medical Solutions USA, Inc.",
      "url": "https://devicerts.com/manufacturers/siemens-medical-solutions-usa-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K143400",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "JAK"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K143400",
      "number_label": null,
      "device_udi_di": null,
      "date": "2015-04-08",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-2440-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-07-25",
      "reason": "SOMATOM Definition Edge, SOMATOM Definition AS, SOMATOM Definition Flash with software syngo.CT  with Software version VA48A_SP5 may result in scanning workflow interruptions and unexpected user notifications. and may result in a delay in diagnosis and/or patient rescans",
      "applies": "this device",
      "via": "K143400"
    },
    {
      "number": "Z-0604-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-10-19",
      "reason": "A problem can occur with the small light marker windows, used for the positioning laser and are integrated in the front cover of the SOMATOM Definition AS and Definition Flash systems, by either becoming loose or potentially dropping out. There is a risk that a person could touch rotating or electrical parts of the gantry if they were to reach through the window opening.",
      "applies": "this device",
      "via": "K143400"
    },
    {
      "number": "Z-1350-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-03-22",
      "reason": "Siemens became aware that the front cover on some Siemens CT system was not properly welded.  It is possible for the welded-in threaded bolt to break out of the ground plate of the front cover if the cover is opened during service.  This could not happen during clinical operation.  Therefore, there is no potential risk for the user or patient and the CT System has no risk for a malfunction during clinical operation",
      "applies": "this device",
      "via": "K143400"
    },
    {
      "number": "Z-1519-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-03-02",
      "reason": "Software bugs in these SOMATOM systems could possibly cause scan aborts, rescans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast medium, accidental x-radiation to the user or impede customer workflow.",
      "applies": "this device",
      "via": "K143400"
    },
    {
      "number": "Z-0021-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-08-14",
      "reason": "software bug issues for SW-Version VA48A_SP0. The following safety issues were resolved: 1)  Correction to improve visual warning and error indication son the gantry display. 2) Correction to improve acquisition data in order to optimize image quality. 3) Correction to improve robustness and general system behavior in some exception handling procedures. 4) Correction to improve auto post processin",
      "applies": "this device",
      "via": "K143400"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K143400",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "JAK",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.1750",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}