Somatom Perspective 10495568
by Siemens Shanghai Medical Equipment Ltd.
Somatom Perspective 10495568 is a Class II device made by Siemens Shanghai Medical Equipment Ltd., cleared under FDA 510(k) K183548 on 15 Apr 2019 (3 earlier submissions) and listed in FDA GUDID since 24 Sep 2016 (UDI-DI 04056869010335), active as of 23 Oct 2026. 5 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK183548 | Date15 Apr 2019 | StatusActive |
| United States | GUDID listing | Number04056869010335 | Date24 Sep 2016 | StatusListed |
Official records
- AccessGUDID record 04056869010335
- FDA 510(k) database K183548
- FDA 510(k) database K151765
- FDA 510(k) database K133590
- FDA 510(k) database K113287
Across registries
- FDA 510(k) K183548 decided 15 Apr 2019 Active
- FDA GUDID UDI-DI 04056869010335 listed 24 Sep 2016 Listed
Recalls
-
- Terminated Z-1448-2018
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure
-
- Terminated Z-2474-2018
- Terminated Z-2475-2018
A potential risk of unnecessary radiation exposure due to a software issue
-
- Terminated Z-0921-2017
Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.
-
- Terminated Z-1646-2015
Software bugs in VC20B SP0a or SP1 software versions may cause issues that could make it necessary to rescan patients. Syngo Main UI may crash if the Patient Browser is scrolled with arrow keys of the keyboard. Tomo images may freeze on the screen when zoomed in or out under the mode of CAREVision. Sporadic displaying error in WorkStream4D application. DB may lock under heavy and multitasks.
Timeline
- Cleared FDA 510(k) K113287
- Cleared FDA 510(k) K133590
- Recall initiated FDA recalls Z-1646-2015
- Cleared FDA 510(k) K151765
- Listed FDA GUDID 04056869010335
- Recall initiated FDA recalls Z-0921-2017
- Recall initiated FDA recalls Z-2474-2018
- Recall initiated FDA recalls Z-2475-2018
- Recall initiated FDA recalls Z-1448-2018
- Cleared FDA 510(k) K183548
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Somatom Perspective
- Generic name
- Full-body CT system
- Model / reference
- 10495568
- Primary DI / UDI-DI
- 04056869010335
- FDA product code
- JAK
- GMDN term
- Full-body CT system
- Device class
- Class II
- Regulation number
- 892.1750
- GUDID distribution status
- In Commercial Distribution
Similar devices in this category
Other devices filed under Full-body CT system.
- MeVis Liver Suite MVLS-1.3 by MEVIS MEDICAL SOLUTIONS AG
- InferOperate® Software 30192180 by Infervision Medical Technology Co., Ltd.
- Revolution by GE HEALTHCARE JAPAN CORPORATION
- NMCT 850 by GE Medical Systems Israel, Functional Imaging
- MeVis Liver Suite by MEVIS MEDICAL SOLUTIONS AG
- StrokeSENS v2.x by Circle Cardiovascular Imaging
- Revolution ACT LMB by GE Hualun Medical Systems Co., Ltd
- Optima CT520 by GE Hangwei Medical Systems Company Ltd.
- Supria by FUJIFILM CORPORATION
- BrainScan CT BrainScan CT 1.3 by BRAINSCAN SPÓŁKA AKCYJNA
Other devices from Siemens Shanghai Medical Equipment Ltd.
See all devices from Siemens Shanghai Medical Equipment Ltd. →