Sophysa Pso-Vtt
Active
510(k)
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number03760409447093 | Date8 Jun 2026 | StatusListed |
| United States | FDA 510(k)SESE | NumberK162108 | Date17 Apr 2017 | StatusActive |
Official records
- AccessGUDID record 03760409447093
- FDA 510(k) database K162108
Recalls
-
- Ongoing Z-1029-2026
Customer complaints of Pressio monitor rebooting.
-
- Ongoing Z-1825-2024
- Ongoing Z-1826-2024
An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Sophysa
- Generic name
- Intracerebral sensor catheter, short-term
- Model / reference
- PSO-VTT
- Primary DI / UDI-DI
- 03760409447093
- 510(k) Number
- FDA product code
- GWM
- GMDN term
- Intracerebral sensor catheter, short-term
- Device class
- Class II
- Regulation number
- 882.1620
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Conditional
Other devices from Sophysa SA
Properties
Other properties (9)
- Sterilise before use: No
- Implantable: No
- Contains latex: No
- Over-the-counter: No
- Convenience kit: No
- Combination product: No
- HCT/P: No
- Premarket exempt: No
- Lot / batch number: No