Spacelabs Arkon Anesthesia Workstation
510(k) K113051
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK113051 | Date22 Mar 2012 | StatusActive |
Recalls
-
- Terminated Z-2589-2018
Arkon Anesthesia Delivery System may go into a failed state (mechanical ventilation ceases) while the machine is in use or while idle.
-
- Terminated Z-0072-2018
Arkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display unit that resulted in a blank screen without audible or visible alarms.
-
- Terminated Z-2093-2017
Reports of the Canister Carrier breaking during absorbent exchange or during anesthesia machine transit, causing a leak in the breathing circuit, which could cause hypoventilation in the patient.
-
- Terminated Z-1460-2017
Spacelabs has received reports of the Arkon Anesthesia Delivery System, Model 99999, shutting down without warning. The power supply has failed in a way that does not trigger the normal on battery alarms.
-
- Terminated Z-1442-2014
The System may go into a controlled failed state if a series of parameters on the user interface are activated in a specific sequence.
-
- Terminated Z-1771-2013
Spacelabs ARKON Anesthesia System is recalled due to software defect. The System Status Computer may incorrectly determine that a communication error has occurred. If this situation occurs, a buzzer will activate and a yellow triangle will be displayed on the System Status Computer screen that alerts the user of this error.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CBK
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.5895
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent