Skip to content

Spacelabs Arkon Anesthesia Workstation

by Spacelabs Healthcare

510(k) K113051

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

6 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK113051 Date22 Mar 2012ClearedStatusActive

Recalls

6 recalls · FDA

    • Terminated Z-2589-2018 FDA Class I initiated 11 Jul 2018

    Arkon Anesthesia Delivery System may go into a failed state (mechanical ventilation ceases) while the machine is in use or while idle.

    • Terminated Z-0072-2018 FDA Class I initiated 11 Oct 2017 on shared 510(k) K113051

    Arkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display unit that resulted in a blank screen without audible or visible alarms.

    • Terminated Z-2093-2017 FDA Class II initiated 14 Apr 2017 on shared 510(k) K113051

    Reports of the Canister Carrier breaking during absorbent exchange or during anesthesia machine transit, causing a leak in the breathing circuit, which could cause hypoventilation in the patient.

    • Terminated Z-1460-2017 FDA Class II initiated 8 Feb 2017

    Spacelabs has received reports of the Arkon Anesthesia Delivery System, Model 99999, shutting down without warning. The power supply has failed in a way that does not trigger the normal on battery alarms.

    • Terminated Z-1442-2014 FDA Class I initiated 10 Mar 2014

    The System may go into a controlled failed state if a series of parameters on the user interface are activated in a specific sequence.

    • Terminated Z-1771-2013 FDA Class II initiated 7 Jun 2013

    Spacelabs ARKON Anesthesia System is recalled due to software defect. The System Status Computer may incorrectly determine that a communication error has occurred. If this situation occurs, a buzzer will activate and a yellow triangle will be displayed on the System Status Computer screen that alerts the user of this error.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
CBK
Device Class
Class II
Regulation
21 CFR Part 868.5895
Advisory Committee
Anesthesiology
FDA Decision
SESE Substantially Equivalent

Other devices from Spacelabs Healthcare Inc.