Spacelabs Healthcare Qube Compact Monitor (91390
510(k) K120616
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK120616 | Date29 Mar 2012 | StatusActive |
Recalls
-
- Ongoing Z-0169-2026
- Ongoing Z-0167-2026
- Ongoing Z-0168-2026
Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.
-
- Terminated Z-1608-2015
The input power can fail even though the power source at the wall has not been interrupted. In addition, there is the potential for the qube Compact Monitor to fail to turn on. This is a retrospective recall related to a recall initiated in 2012 (Z-0403-2013).
-
- Terminated Z-1145-2015
Spacelabs Healthcare qube Bedside Monitors, Model 91390, are recalled because the firm has received multiple reports of qube monitors failing to boot up or returning to factory default configuration settings following power on or reset.
-
- Terminated Z-1852-2014
Spacelabs Healthcare is recalling Qube Compact Monitor, Model 91390, due to a potential chance of losing network connection after switching from hardwired Ethernet connection to using the Qube Docking Station without powering down the device. The bedside monitor and all of its alarms will continue to function normally. No one has been injured as a result of this issue.
-
- Terminated Z-0351-2014
Spacelabs Healthcare qube Compact Patient Monitor, Model 91390, may fail to power ON due to incorrect programmed microcontroller component.
-
- Terminated Z-0188-2014
Spacelabs Healthcare elance Vital Signs Monitor with Option S, model 93330, is recalled because the monitor will not permit connection of any of the Spacelabs TruLink line of SpO2 sensors.
-
- Terminated Z-0403-2013
Spacelabs Healthcare has learned through several reports that the input circuit may fail and the monitor will go on battery mode. The monitor will function normally until the batteries become depleted and then the monitor will shut down. In addition the monitor may fail to turn on due to an internal short.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MHX
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.1025
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent