SpaceStation MRI
by B. Braun Melsugen AG 510(k)
Active
510(k) K191910
Regulatory status
Recall ongoing on a related submission. 7 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK191910 | Date12 Mar 2020 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-2972-2024
Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms. Additional devices have been affected.
-
- Ongoing Z-2437-2024
- Ongoing Z-2433-2024
- Ongoing Z-2434-2024
- Ongoing Z-2435-2024
Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
-
- Ongoing Z-0123-2024
- Ongoing Z-0124-2024
Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to prevent permanent impairment to body structures or body functions. In some cases, this may be life-threatening or may lead to death.
-
- Terminated Z-0819-2022
Units not properly inspected prior to release to the market, may not function as indicated in the product labeling, resulting in syringe detection and infusion accuracy that can result in delays in therapy, overinfusion, or underinfusion.
Specification
- 510(k) Number
- FDA Product Code
- MRZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5725
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent