Speedlink Transverse Connector
510(k) K002082
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK002082 | Date6 Oct 2000 | StatusActive |
Recalls
-
- Terminated Z-0525-2010
The titanium stock used to manufacturer the product is reported to exhibit areas of alloy segregation, which could result in a weakening of the strength of the metal.
-
- Terminated Z-0027-2011
- Terminated Z-0026-2011
- Terminated Z-0029-2011
- Terminated Z-0028-2011
- Terminated Z-0033-2011
- Terminated Z-0030-2011
- Terminated Z-0032-2011
The firm received complaints that the device was damaged or fractured while the surgeon was bending it prior to implantation. There were also complaints that the cam may dislodge during tightening. The firm added additional instructions for use to reduce the probability of these two types of events.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- KWP
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3050
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent