  "format": "devicerts.device.v1",
  "cite_as": "SPiN Thoracic Navigation System, Veran Medical Technologies Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/spin-thoracic-navigation-system-jak (data as of 2026-11-23).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/spin-thoracic-navigation-system-jak",
    "json_url": "https://devicerts.com/devices/spin-thoracic-navigation-system-jak.json",
    "markdown_url": "https://devicerts.com/devices/spin-thoracic-navigation-system-jak.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-23",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/spin-thoracic-navigation-system-jak",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "SPiN Thoracic Navigation System",
    "registry_name": "SPiN Thoracic Navigation System",
    "url": "https://devicerts.com/devices/spin-thoracic-navigation-system-jak",
    "manufacturer": {
      "name": "Veran Medical Technologies Inc.",
      "url": "https://devicerts.com/manufacturers/veran-medical-technologies-inc",
      "country": "United States",
      "country_basis": "FDA 510(k) applicant",
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K170023",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "JAK"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K170023",
      "number_label": null,
      "device_udi_di": null,
      "date": "2017-05-15",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-2663-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-08-10",
      "reason": "Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments)  manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-2661-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-08-10",
      "reason": "Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments)  manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-2660-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-08-10",
      "reason": "Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments)  manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-2659-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-08-10",
      "reason": "Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments)  manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-2662-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-08-10",
      "reason": "Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments)  manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-2664-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-08-10",
      "reason": "Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments)  manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-2269-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-05-15",
      "reason": "Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-2262-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-05-15",
      "reason": "Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-2268-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-05-15",
      "reason": "Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-2263-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-05-15",
      "reason": "Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-2267-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-05-15",
      "reason": "Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-2264-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-05-15",
      "reason": "Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-2266-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-05-15",
      "reason": "Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-2265-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-05-15",
      "reason": "Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay",
      "applies": "this device",
      "via": "K170023"
    },
    {
      "number": "Z-1103-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2022-12-13",
      "reason": "A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged  procedure/surgery due to additional time to  troubleshoot the device",
      "applies": "this device",
      "via": "K170023"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K170023",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "JAK",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.1750",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}