StealthStation S8 9735667
by Medtronic Navigation Inc. 2 variants
Ref 9735667
StealthStation S8 9735667 is a multi-purpose stereotactic surgery system (EU MDR Class I; FDA Class II; FDA product code HAW, Neurological stereotaxic Instrument) made by Medtronic Navigation Inc., registered in EUDAMED on 25 Sep 2026 (Basic UDI-DI 0763000B00004737V), cleared under FDA 510(k) K162309 on 31 Mar 2017 and listed in FDA GUDID since 18 Jun 2021 (UDI-DI 00763000359812), active as of 8 Nov 2026. 2 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
1 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number0763000B00004737V 00763000359812 | Date25 Sep 2026 | StatusActive |
| United States | GUDID listing GUDID | Number00763000359812 | Date18 Jun 2021 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK162309 | Date31 Mar 2017 | StatusActive |
Official records
- EUDAMED public site 0763000B00004737V
- AccessGUDID record 00763000359812
- FDA 510(k) database K162309
From the FDA record
- Device description
- MAIN CART 9735667 STEALTH S8 PLUS
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- EUDAMED Basic UDI-DI 0763000B00004737V registered 25 Sep 2026 Active
- FDA 510(k) K162309 decided 31 Mar 2017 Active
- FDA GUDID UDI-DI 00763000359812 listed 18 Jun 2021 Listed
Recalls
-
- Ongoing Z-1776-2024
Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.
-
- Terminated Z-1995-2020
Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the lead placement and subsequent imaging).
Timeline
- Cleared FDA 510(k) K162309
- Recall initiated FDA recalls Z-1995-2020
- Listed FDA GUDID 00763000359812
- Recall initiated FDA recalls Z-1776-2024
- Registered EUDAMED 0763000B00004737V
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- StealthStation S8
- Generic name
- Multi-purpose stereotactic surgery system
- Model / reference
- 9735667
- Primary DI / UDI-DI
- 00763000359812
- Basic UDI-DI
- 0763000B00004737V
- 510(k) Number
- FDA product code
- HAW
- EMDN code
- Z12011401 Surgical navigation system
- GMDN term
- Multi-purpose stereotactic surgery system
- Device class
- Class I EUDAMEDClass II GUDID
- Regulation number
- 882.4560
- EU registration type
- Standard device
- EU market of registration
- Netherlands
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device class | Class I | Class II |
Similar devices in this category
Other devices filed under SURGICAL NAVIGATION SYSTEM.
- TGS Navigation 8000-010-003 by Stryker Leibinger
- Joimax® Intracs® EM Patient Mapper - AP JINTPM2245-AP by Joimax GmbH
- RIWOtrack Flexsensor N 01 2950 by Fiagon GmbH
- Tracer Set 1001 30-199 by Ortoma AB (Publ)
- Joimax® Intracs® EM Sensor Clip JINTSC2205 by Joimax GmbH
- FlexPointer 1.5 mm by Fiagon GmbH
- Navient 955-NC-NC by Claronav Kolahi
- ARROW IPN916842 RHY-100-ECGB-1 by TELEFLEX INCORPORATED
- ARVIS IN-50000 by KICO KNEE INNOVATION COMPANY PTY LIMITED
- IGI AIG0005-V2 by Image Navigation