# StealthStation System with Cranial Software

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/stealthstation-system-with-cranial-software-haw>
- Manufacturer: [Medtronic Navigation Inc.](https://devicerts.com/manufacturers/medtronic-navigation-inc) (United States)
- Device class: Class II (US FDA)
- FDA product code: HAW

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K153660 · decision 2016-09-14 · decision SESE · status: Active

## Recalls

- Z-1666-2023 · Class I · ongoing · initiated 2023-04-11: In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury
- Z-1995-2020 · Class I · terminated · initiated 2019-08-09: Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the lead placement and subsequent imaging).
- Z-0552-2019 · Class I · terminated · initiated 2018-09-26: An incorrect display of the Biopsy Depth Gauge graphical representation in the Guidance View using Cranial software, which resulted in the biopsy of healthy tissue.

## Identifiers

- 510(k) Number: K153660 [FDA 510(k)]
- FDA product code: HAW [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 882.4560 [FDA 510(k)]
- Advisory committee: Neurology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: StealthStation System with Cranial Software, Medtronic Navigation Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/stealthstation-system-with-cranial-software-haw (data as of 2026-11-24).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-11-08
- Statuses computed as of: 2026-11-24
- JSON: <https://devicerts.com/devices/stealthstation-system-with-cranial-software-haw.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.