Skip to content

Stimulus Switching Pod SP2

by Natus Neurology Incorporated

Ref 515-020500

Active European Union. Also listed in GUDID (US).

CE MDR

Valid until

Not listed

EUDAMED holds no validity date

FDA recalls

None ongoing

4 closed

Documents

None linked in EUDAMED

A group certificate may exist

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source EUDAMED, GUDID, synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDR EUDAMED Number038283NA00151E9 Basic UDI-DI 00382830037020UDI-DIDate15 Apr 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number00382830037020 UDI-DI Date19 Sep 2016ListedStatusListed GUDID: in commercial distribution

Certificates

0 certificates · EUDAMED

EUDAMED lists no certificate for this device's Basic UDI-DI.

A certificate issued for a device group may exist without being linked here.

Recalls

4 recalls · FDA

    • Terminated Z-1256-2010 FDA Unclassified (pre-2012) initiated 8 Mar 2010 on shared 510(k) K991054

    CareFusion NeuroCare is voluntarily implementing a field correction to delete certain NicoletOne software protocols, which users may interpret incorrectly when conducting patient evaluations. This may result in the misinterpretation of the left versus right side of the brain. Users are being provided with instructions for removing these protocols and verifying that the deletion has been prop

    • Terminated Z-2459-2008 FDA Unclassified (pre-2012) initiated 13 Jun 2008 on shared 510(k) K991054

    NicoletOne 5.30.2 Software when used with CSeries amplifiers in Fixed Room LTM Configurations with patient room kits. Cardinal Health NeuroCare has confirmed that the 3rd party room relay connected to the A2 wall plate does not always activate if the patient event button connected to the CSeries Amplifier is pressed twice within one second.

    • Terminated Z-2389-2008 FDA Unclassified (pre-2012) initiated 5 Jun 2008 on shared 510(k) K991054

    Cardinal Health NeuroCare is implementing a voluntary field correction to replace a quantity of photic adapter cables, part number 085-463700, used to connect the NicLED Photic Strobe option, part number 698-625500, with NicoletOne EEG system. The photic adapter cable has a wiring error which delivers a reduced voltage and results in a decreased intensity of the NicLED photic. This reduced intensi

    • Terminated Z-0380-2007 FDA Unclassified (pre-2012) initiated 25 Aug 2006 on shared 510(k) K991054

    A software anomaly exists in the 5.20.1038 NicoletOne LTM, Sleep, nEEG, vEEG and ICU Monitor systems using the M, and/or C series amplifiers. If this anomaly occurs, the system will display the data for channel one in all channels on NicoletOne LTM, Sleep, nEEG, vEEG or ICU Monitor systems.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDR Active
  • United States GUDID listing Listed

Specification

EUDAMED · GUDID

Trade Name
Stimulus Switching Pod SP2 EUDAMED
Natus Neurology GUDID
Generic Name
Electromyograph
Device Name
Nicolet EDX
Model / Reference
515-020500
Primary DI / UDI-DI
00382830037020
Basic UDI-DI
038283NA00151E9
Direct Marketing DI
00382830037020
510(k) Number
FDA Product Code
GWF
EMDN Code
Z12100480 Electromyography instruments - hardware accessories
GMDN Term
Electromyograph
Device Class
Class IIb EUDAMED
Class II GUDID
Regulation Number
882.1870
EU market of registration
Ireland
GUDID distribution status
In Commercial Distribution

Truth per registry

2 sources · 1 field differs

FieldEUDAMEDGUDID
Device ClassEUDAMED
Class IIb
GUDID
Class II

Manufacturer details

EUDAMED SRN
US-MF-000004246
Authorised representative
Natus Manufacturing Ltd IE-AR-000000884

Other devices from Natus Neurology Incorporated

See all devices from Natus Neurology Incorporated →

Properties

  • Prescription Use
  • Serial Number
  • Manufacturing Date

Other properties (21)

  • Sterile: No
  • Single Use: No
  • Implantable: No
  • Reusable: No
  • Contains Latex: No
  • Reprocessed: No
  • CMR Substance: No
  • Human-derived: No
  • Human Tissues: No
  • Animal Tissues: No
  • Endocrine Disruptor: No
  • Administering Medicine: No
  • Measuring Function: No
  • Incorporates Medicinal Product: No
  • Over-the-Counter: No
  • Convenience Kit: No
  • Combination Product: No
  • HCT/P: No
  • Premarket Exempt: No
  • Lot / Batch Number: No
  • Expiration Date: No