Stockert Heater-Cooler System 3T

by Sorin Group Deutschland GmbH

510(k) K052601

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

18 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 10 Oct 2026. What this means

Regulatory status

as of registry sync 10 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK052601 Date6 Jun 2006ClearedStatusActive

Recalls

18 recalls · FDA

    • Terminated Z-1498-2020 FDA Class II initiated 24 Feb 2020

    The firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM contamination.

    • Terminated Z-0220-2020 FDA Class II initiated 1 Nov 2018

    The firm has become aware that due to a chemical reaction, the disinfectant/water preservative used to disinfect the device to inhibit/limit the growth micro-organisms, may be rendered ineffective. In addition, the firm is releasing a design upgrade to reduce the risk of potential emission of aerosols from the 3T.

    • Terminated Z-1099-2019 FDA Class II initiated 14 Sep 2018

    For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).

    • Terminated Z-1901-2018 FDA Class II initiated 20 Apr 2018

    Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm.

  • 6 recalls, same reason

    • Terminated Z-2774-2017 FDA Class II initiated 30 Jun 2017
    • Terminated Z-2776-2017 FDA Class II initiated 30 Jun 2017
    • Terminated Z-2775-2017 FDA Class II initiated 30 Jun 2017
    • Terminated Z-2773-2017 FDA Class II initiated 30 Jun 2017
    • Terminated Z-2772-2017 FDA Class II initiated 30 Jun 2017
    • Terminated Z-2771-2017 FDA Class II initiated 30 Jun 2017

    Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.

  • 6 recalls, same reason

    • Terminated Z-2080-2015 FDA Class II initiated 15 Jun 2015
    • Terminated Z-2081-2015 FDA Class II initiated 15 Jun 2015
    • Terminated Z-2079-2015 FDA Class II initiated 15 Jun 2015
    • Terminated Z-2076-2015 FDA Class II initiated 15 Jun 2015
    • Terminated Z-2078-2015 FDA Class II initiated 15 Jun 2015
    • Terminated Z-2077-2015 FDA Class II initiated 15 Jun 2015

    Potential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.

    • Terminated Z-0682-2017 FDA Class II initiated 1 Aug 2014

    On August 1, 2014 Sorin Group USA, Inc. informed consignees via letter to review the disinfection and monitoring practices and to strictly adhere to the instructions in the Instructions for Use (IFU).

    • Terminated Z-1186-2016 FDA Class II initiated 15 Aug 2012

    On August 15, 2012, Sorin informed consignees via letter of changes made to create Version 12 of the Heater-Cooler 3T IFU. The revised IFU included enhancements to achieve water quality that meets the standard for German drinking water.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
DWC
Device Class
Class II
Regulation
21 CFR Part 870.4250
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent

Other devices from Sorin Group Deutschland GmbH

See all devices from Sorin Group Deutschland GmbH →