Stockert Heater-Cooler System 3T
by Sorin Group Deutschland GmbH
510(k) K052601
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK052601 | Date6 Jun 2006 | StatusActive |
Recalls
-
- Terminated Z-1498-2020
The firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM contamination.
-
- Terminated Z-0220-2020
The firm has become aware that due to a chemical reaction, the disinfectant/water preservative used to disinfect the device to inhibit/limit the growth micro-organisms, may be rendered ineffective. In addition, the firm is releasing a design upgrade to reduce the risk of potential emission of aerosols from the 3T.
-
- Terminated Z-1099-2019
For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).
-
- Terminated Z-1901-2018
Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm.
-
- Terminated Z-2774-2017
- Terminated Z-2776-2017
- Terminated Z-2775-2017
- Terminated Z-2773-2017
- Terminated Z-2772-2017
- Terminated Z-2771-2017
Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
-
- Terminated Z-2080-2015
- Terminated Z-2081-2015
- Terminated Z-2079-2015
- Terminated Z-2076-2015
- Terminated Z-2078-2015
- Terminated Z-2077-2015
Potential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.
-
- Terminated Z-0682-2017
On August 1, 2014 Sorin Group USA, Inc. informed consignees via letter to review the disinfection and monitoring practices and to strictly adhere to the instructions in the Instructions for Use (IFU).
-
- Terminated Z-1186-2016
On August 15, 2012, Sorin informed consignees via letter of changes made to create Version 12 of the Heater-Cooler 3T IFU. The revised IFU included enhancements to achieve water quality that meets the standard for German drinking water.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DWC
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.4250
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent