Skip to content

Stratus Cs Acute Care Nt-Probnp Testpak, Pbnp Calpak Calibrator, Pbnp Dilpak Diluent

by Dade Behring, Inc.

510(k) K043476

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 11 Oct 2026. What this means

Regulatory status

as of registry sync 11 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK043476 Date15 Feb 2005ClearedStatusActive

Recalls

2 recalls · FDA

    • Terminated Z-0879-06 FDA Unclassified (pre-2012) initiated 3 Apr 2006

    Foil sealing of the TestPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the magnitude of the inaccuracy may vary based on the degree of occlusion.

    • Terminated Z-0885-06 FDA Unclassified (pre-2012) initiated 3 Apr 2006

    Foil sealing of the DilPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the magnitude of the inaccuracy may vary based on the degree of occlusion.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
NBC
Device Class
Class II
Regulation
21 CFR Part 862.1117
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent

Other devices from Dade Behring, Inc.

See all devices from Dade Behring, Inc. →