Skip to content

Stratus Cs D-Dimer

by Dade Behring, Inc. 510(k)

Active United States

510(k) K051597

Regulatory status

as of 28 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK051597 Date17 Aug 2005ClearedStatusActive

Coverage

1 jurisdiction

United States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of AmericaUnited States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not known
  • Not covered

United States — FDA 510(k) Active

Recalls

1 recall · FDA

    • Terminated Z-0877-06 FDA Unclassified (pre-2012) initiated 3 Apr 2006

    Foil sealing of the TestPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the magnitude of the inaccuracy may vary based on the degree of occlusion.

Specification

510(k)

510(k) Number
FDA Product Code
DAP
Device Class
Class II
Regulation
21 CFR Part 864.7320
Advisory Committee
Hematology
FDA Decision
SESE Substantially Equivalent

Other devices from Dade Behring, Inc.

See all devices from Dade Behring, Inc. →