# Stryker Navigation System - Hip Module

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/stryker-navigation-system-hip-module-haw>
- Manufacturer: [Stryker Instruments](https://devicerts.com/manufacturers/stryker-instruments) (United States)
- Name in the registry: STRYKER NAVIGATION SYSTEM - HIP MODULE
- Device class: Class II (US FDA)
- FDA product code: HAW

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K022365 · decision 2003-01-22 · decision SESE · status: Active

## Recalls

- Z-1810-2015 · Class II · terminated · initiated 2015-05-04: Stryker initiated a device recall of the Navigation System II-Cart, Camera Articulated Arm because there is a potential that the weld seam between the main stud and the two flaps on the Articulated Arm was not welded in the correct location, which may result in a potential failure of the arm joint.
- Z-2288-2008 ·  · terminated · initiated 2008-06-05: Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.
- Z-2289-2008 ·  · terminated · initiated 2008-06-05: Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.
- Z-2290-2008 ·  · terminated · initiated 2008-06-05: Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.
- Z-2291-2008 ·  · terminated · initiated 2008-06-05: Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.
- Z-2292-2008 ·  · terminated · initiated 2008-06-05: Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.
- Z-2293-2008 ·  · terminated · initiated 2008-06-05: Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.
- Z-2294-2008 ·  · terminated · initiated 2008-06-05: Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.
- Z-0020-05 ·  · terminated · initiated 2004-07-14: The tracker is not locking adequately to the anchoring pin, resulting in unacceptable movement, which will affected the patient tracker system.
- Z-0021-05 ·  · terminated · initiated 2004-07-14: The tracker is not locking adequately to the anchoring pin, resulting in unacceptable movement, which will affected the patient tracker system.

## Identifiers

- 510(k) Number: K022365 [FDA 510(k)]
- FDA product code: HAW [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 882.4560 [FDA 510(k)]
- Advisory committee: Neurology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Stryker Navigation System - Hip Module, Stryker Instruments. deviCERTS medical device catalogue. https://devicerts.com/devices/stryker-navigation-system-hip-module-haw (data as of 2026-11-24).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-24
- JSON: <https://devicerts.com/devices/stryker-navigation-system-hip-module-haw.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/stryker-navigation-system-hip-module-haw
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.