  "format": "devicerts.device.v1",
  "cite_as": "Superion Indirect Decompression System (IDS), Boston Scientific Neuromodulation. deviCERTS medical device catalogue. https://devicerts.com/devices/superion-interspinous-spacer-p140004 (data as of 2026-11-26).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/superion-interspinous-spacer-p140004",
    "json_url": "https://devicerts.com/devices/superion-interspinous-spacer-p140004.json",
    "markdown_url": "https://devicerts.com/devices/superion-interspinous-spacer-p140004.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA PMA",
        "url": "https://open.fda.gov/apis/device/pma/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-26",
    "status_as_of": "2026-11-26",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/superion-interspinous-spacer-p140004",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Superion Indirect Decompression System (IDS)",
    "registry_name": "Superion™ Indirect Decompression System (IDS)",
    "url": "https://devicerts.com/devices/superion-interspinous-spacer-p140004",
    "manufacturer": {
      "name": "Boston Scientific Neuromodulation",
      "url": "https://devicerts.com/manufacturers/boston-scientific-neuromodulation",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class III",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "PMA number",
        "values": [
          {
            "value": "P140004",
            "registries": [
              "FDA PMA"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "NQO"
    ],
    "emdn": [],
    "gmdn": [],
    "description": {
      "text": "The Superion Indirect Decompression System (IDS) is an interspinous spacer/plate prosthesis designed to relieve neurogenic intermittent claudication caused by moderate degenerative lumbar spinal stenosis. This device is intended for skeletally mature patients with confirmed radiographic evidence of ligamentum flavum thickening, lateral recess or central canal/foraminal narrowing, and associated nerve root impingement or thecal sac compression.\n\nThe Superion ISS is implanted at one or two adjacent lumbar levels (L1–L5) in patients with impaired physical function, as defined by a Zurich Claudication Questionnaire score ≥ 2.0 and the ability to sit for 50 minutes without pain. Supplied sterile and intended for single-use, the device carries FDA PMA approval (P140004) and is classified as a Class III medical device under the NQO product code. Storage and handling must comply with sterile instrument protocols.",
      "ai_generated": true,
      "generated_on": "2026-09-18",
      "note": "Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data is the record."
    },
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "PMA",
      "region": "United States",
      "type": "FDA PMA",
      "number": "P140004",
      "number_label": null,
      "device_udi_di": null,
      "date": "2015-05-20",
      "date_kind": "decision",
      "decision_code": "APPR",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-2485-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-06-06",
      "reason": "As a result of a comprehensive product performance review it was determine that the device and instruments are not consistently meeting expected performance levels.  Issues include breakage of the instruments and implant/device.",
      "applies": "this device",
      "via": "P140004"
    },
    {
      "number": "Z-1972-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-05-12",
      "reason": "Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.",
      "applies": "this device",
      "via": "P140004"
    },
    {
      "number": "Z-0230-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2022-10-12",
      "reason": "Due to Indirect Decompression System devices exhibiting a potential for weld breakage at the spindle cap-to-main body interface.  As a result of the separation of the spindle cap from the main body it may prevent proper deployment of the device during implantation.",
      "applies": "this device",
      "via": "P140004"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identity",
      "fields": [
        {
          "label": "Trade name",
          "values": [
            {
              "value": "SUPERION INTERSPINOUS SPACER",
              "registries": [
                "FDA PMA"
              ]
            }
          ]
        },
        {
          "label": "Generic name",
          "values": [
            {
              "value": "Prosthesis, spinous process spacer/plate",
              "registries": [
                "FDA PMA"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "PMA number",
          "values": [
            {
              "value": "P140004",
              "registries": [
                "FDA PMA"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "NQO",
              "registries": [
                "FDA PMA"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class III",
              "registries": [
                "FDA PMA"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Orthopedic",
              "registries": [
                "FDA PMA"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "APPR",
              "registries": [
                "FDA PMA"
              ]
            }
          ]
        }
      ]
    }
  ]
}