Surfix 285312SND
by Newdeal Sas
Surfix 285312SND is an orthopaedic bone screw (FDA Class II; FDA product code HRS, Plate, Fixation, Bone) made by Newdeal Sas, cleared under FDA 510(k) K093914 on 14 Apr 2011 (5 earlier submissions) and listed in FDA GUDID since 9 Oct 2017 (UDI-DI 10381780048893), active as of 23 Oct 2026. 9 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10381780048893 | Date9 Oct 2017 | StatusListed |
| United States | FDA 510(k)SESE | NumberK093914 | Date14 Apr 2011 | StatusActive |
Official records
- AccessGUDID record 10381780048893
- FDA 510(k) database K093914
- FDA 510(k) database K091609
- FDA 510(k) database K073375
- FDA 510(k) database K070447
- FDA 510(k) database K063820
- FDA 510(k) database K070241
From the FDA record
- Device description
- Titanium alloy Newdeal sterile locking system fixation consists of screws and washers. This system is available in different sizes. The osteosynthesis screws are fixed to the bone through the plate holes, creating a single implant/screw unit. The system is locked by means of washers drilled into the threaded lipped sockets at the top of each hole, thus blocking each screw head.
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- Labelled as not made with natural rubber latex
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K093914 decided 14 Apr 2011 Active
- FDA GUDID UDI-DI 10381780048893 listed 9 Oct 2017 Listed
Recalls
-
- Ongoing Z-0548-2022
The affected screwdrivers are out of specification and may not mate with the Surfix Alpha Lock screw
-
- Terminated Z-2443-2018
- Terminated Z-2444-2018
- Terminated Z-2445-2018
- Terminated Z-2446-2018
- Terminated Z-2447-2018
Use of the impacted product may cause a superficial infection requiring PO antibiotics and wound care or deep infection requiring IV antibiotics and device removal.
-
- Terminated Z-1697-2014
New Deal has identified through a complaint report that some 3.5 mm Diameter x 22 mm Length Surfix screws in Lot no. F626 (Catalogue No. 285322S) have been packaged and labeled as 3.5mm Diameter x 30 mm Length Surfix screws (Catalogue No. 285330S).
-
- Terminated Z-2123-2012
Incorrect technique applied may cause potential rupture of the tibialis anterior tendon during surgery.
-
- Terminated Z-0955-2011
Integra LifeSciences Corporation has decided to initiate a voluntary recall of the Uni-Comp Kits due to inaccurate information relative to the recommended sterilization parameters described in the products Instructions for Use (IFU). The sterilization parameters noted in the Uni-Comp IFU were incorrect. It is noted that the sterilization recommendations in the UNI-Comp IFU in which the Exposure
-
- Terminated Z-0307-2009
Reports of the Uni-CP Compression Forceps breaking during the Compression of the Uni-CP plate
Timeline
- Cleared FDA 510(k) K093914
- Recall initiated FDA recalls Z-2123-2012
- Recall initiated FDA recalls Z-1697-2014
- Listed FDA GUDID 10381780048893
- Recall initiated FDA recalls Z-2443-2018
- Recall initiated FDA recalls Z-2444-2018
- Recall initiated FDA recalls Z-2445-2018
- Recall initiated FDA recalls Z-2446-2018
- Recall initiated FDA recalls Z-2447-2018
- Recall initiated FDA recalls Z-0548-2022
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Surfix
- Generic name
- Orthopaedic bone screw, non-bioabsorbable, sterile
- Model / Ref / Catalog
- 285312SND
- Primary DI / UDI-DI
- 10381780048893
- FDA product code
- HRS
- GMDN term
- Orthopaedic bone screw, non-bioabsorbable, sterile
- Device class
- Class II
- Regulation number
- 888.3030
- GUDID distribution status
- In Commercial Distribution
Similar devices in this category
Other devices filed under Orthopaedic bone screw, non-bioabsorbable, sterile.
- ACUMED 30-0244-S by ACUMED LLC
- Cannulated Screw 110009804 by Biomet Orthopedics, Inc.
- Cannulated Screw 110009886 by Biomet Orthopedics, Inc.
- CHIMAERA Hip Fracture System 1 99-T93520 by ORTHOFIX SRL
- OsteoMed 330-2020-SP-05 by Osteomed Corp.
- ACUMED 30-0282-S by ACUMED LLC
- Centronail Titanium Humeral Nailing System 1 99-T786025 by ORTHOFIX SRL
- Unity Total Knee System 112.054.92 by CORIN LTD
- OsteoMed 337-2716-SP-05 by Osteomed Corp.
- PAO Cortical Screw 05.07.0035 by Medacta International SA