Surgical Cutter
by Promex, Inc.
510(k) K955168
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK955168 | Date22 Jan 1996 | StatusActive |
Recalls
-
- Terminated Z-2767-2015
Two cables within the console were inadvertently twisted during manufacturing. The individual wires within the cables may break and/or contact one another where the cables terminate into two connectors. When these wires break or contact one another (i.e., short), various types of malfunction may occur with the console. One type of malfunction may result in unexpected aspiration increase.
-
- Terminated Z-2045-2016
This recall has been initiated due to a defective component within the handpiece which will cause the handpiece to fail. The failure will result in the inability to cut tissue with the handpiece, rendering it non-functional. Should this failure occur, a replacement Myriad handpiece will be required. If a replacement handpiece is not available, alternate instrumentation for tissue removal will be necessary.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- GEI
- Device Class
- Class II
- Regulation
- 21 CFR Part 878.4400
- Advisory Committee
- General, Plastic Surgery
- FDA Decision
- SESE Substantially Equivalent