Surgicot Steam Test Pack
510(k) K854902
Active
Valid until
No expiry
FDA recalls
5 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK854902 | Date20 May 1986 | StatusActive |
Recalls
-
- Ongoing Z-0074-2027
- Ongoing Z-0071-2027
- Ongoing Z-0070-2027
- Ongoing Z-0075-2027
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
-
- Ongoing Z-1793-2022
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
-
- Terminated Z-0605-2014
STERIS has identified that the population of the G. stearothermophilus spores that monitor STEAM sterilization cycles does not meet specifications for Lot # 131004 of the Verify Dual Species Self-Contained Biological Indicators. As a result the spore count on the affected product is lower than the labeled minimum and the product may incorrectly indicate that a 5-log spore kill was achieved.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- FRC
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.2800
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent