> Active: European Union. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/symbia_t16-10275010>
- Manufacturer: [Siemens Medical Solutions USA, Inc.](https://devicerts.com/manufacturers/siemens-medical-solutions-usa-inc) (United States)
- Device class: Class IIb (EU MDR); Class II (US FDA)
- FDA product code: KPS
- EMDN: Z11020201 CT/GAMMA CAMERA SYSTEMS
- GMDN: SPECT/CT system (45016)

## Registrations and authorizations

- **EUDAMED** · CE/MDR · European Union · Basic UDI-DI 0405686900135UT · UDI-DI 04056869001203 · registered 2026-06-19 · status: Active
- **GUDID** · GUDID listing · United States · UDI-DI 04056869001203 · listed 2016-09-24 · status: Listed (GUDID: in commercial distribution)

## Recalls

- Z-2934-2020 · Class II · terminated · initiated 2020-08-14 · the recall lists this device's catalog number (version and lot not checked): This letter is to inform you of the possibility that your Symbia S or Symbia T system(s) may have unintended slow motion of the detectors in specific situations which could pose a potential risk.
- Z-2436-2015 · Class II · terminated · initiated 2015-07-31 · the recall lists this device's catalog number (version and lot not checked): Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbia lntevo systems. These fuses, in conjunction with a specific failure mode of the Power Controller assembly, could lead to the overheating of the assembly.
- Z-0912-2015 · Class II · terminated · initiated 2014-08-12 · the recall lists this device's catalog number (version and lot not checked): Siemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression related injuries.
- Z-1649-2009 ·  · terminated · initiated 2009-03-11 · the recall lists this device's catalog number (version and lot not checked): Patient injury can occur when the patient is lying on the patient bed and the automatic collimator exchanger is undergoing a collimator change operation at the launch of an acquisition workflow. Detector motions required for this change can cause collision with the patient or pallet if it is in brain scan position, i.e. the patient or pallet is into the field of view.
- Z-0190-2009 ·  · terminated · initiated 2008-09-29 · the recall lists this device's catalog number (version and lot not checked): The protective plastic cap over the CT gantry power switch on the Line Connection Box may come loose, exposing the energized electrical contacts within the switch, thereby causing an electric hazard.

## Identity

- Trade name: Symbia\_T16 [EUDAMED, FDA GUDID (AccessGUDID)]
- Generic name: SPECT/CT system [FDA GUDID (AccessGUDID)]
- Device name: SPECT-CT\_10\_IIb\_HES [EUDAMED]
- Model / reference: 10275010 [EUDAMED, FDA GUDID (AccessGUDID)]; SPECT-CT\_10\_IIb\_HES [EUDAMED]

## Identifiers

- Primary DI / UDI-DI: 04056869001203 [EUDAMED, FDA GUDID (AccessGUDID)]
- Basic UDI-DI: 0405686900135UT [EUDAMED]
- FDA product code: KPS [FDA GUDID (AccessGUDID)]
- EMDN code: Z11020201 [EUDAMED]
- GMDN term: SPECT/CT system (45016) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class IIb [EUDAMED]; Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 892.1200 [FDA GUDID (AccessGUDID)]
- EU registration type: Standard device [EUDAMED]
- EU market of registration: Italy [EUDAMED]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Serial number: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Sterilise before use: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reusable: No [EUDAMED]
- Contains latex: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reprocessed single-use device: No [EUDAMED]
- Contains CMR substances (carcinogenic, mutagenic or toxic to reproduction): No [EUDAMED]
- Incorporates a substance derived from human blood or plasma: No [EUDAMED]
- Contains human tissues or cells: No [EUDAMED]
- Contains animal tissues or cells: No [EUDAMED]
- Contains endocrine-disrupting substances: No [EUDAMED]
- Administering medicine: No [EUDAMED]
- Measuring function: No [EUDAMED]
- Incorporates a medicinal substance: No [EUDAMED]
- Prescription use: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Lot / batch number: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Symbia\_T16 10275010, Siemens Medical Solutions USA, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/symbia\_t16-10275010 (data as of 2026-11-27).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [EUDAMED](https://ec.europa.eu/tools/eudamed), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-24
- Statuses computed as of: 2026-11-27
- JSON: <https://devicerts.com/devices/symbia_t16-10275010.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/symbia\_t16-10275010
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.