Symphony Glucose Management System
by Animas Corp. 510(k)
Active
510(k) K080639
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK080639 | Date24 Jun 2008 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1376-2012
There is an error in the pump software where the OneTouch(R) Ping(R) insulin pump did not allow the user to save changes to the time and date on the pump if those changes were made on February 29, 2012 (Leap Day). Users who did not attempt to change the time or date on their pump on February 29, 2012 are not affected by this issue.
-
- Terminated Z-2394-2010
sensor could be damaged
-
- Terminated Z-1560-2010
Software malfunction which disallows users to download, view and print information from their pump (including blood glucose, insulin delivery, carbohydrate and pump events). Additionally, users are unable to customize the built-in food database and pump settings using the software.
-
- Terminated Z-0955-2009
incorrect dosing recommendation
-
- Terminated Z-0291-2009
- Terminated Z-0290-2009
Unintentional rebooting: Pump products exhibit an intermittent loss of power due intermittent loss of contact between battery cap and battery canister resulting in the device resetting. The failure of the battery cap may result in failure of the device to administer insulin therapy which may result in hyperglycemia.
-
- Terminated Z-0283-2009
display old, inaccurate values for bolus amount delivered and amount of planned bolus totals
Specification
- 510(k) Number
- FDA Product Code
- LZG
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5725
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent