Synapse PACS
510(k) K190232
Synapse PACS is a medical device (FDA Class II; FDA product code LLZ, System, Image Processing, Radiological) made by Fujifilm Corporation, cleared under FDA 510(k) K190232 on 26 Mar 2019, active as of 22 Oct 2026. 11 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
9 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK190232 | Date26 Mar 2019 | StatusActive |
Official records
- FDA 510(k) database K190232
From the FDA record
- Received by the FDA
- 6 Feb 2019
- FDA decision
- 26 Mar 2019 (Substantially Equivalent)
- FDA review time
- 48 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-3030-2026
Software issue may cause the ruler measurement to be inaccurate.
-
- Ongoing Z-1407-2025
The incorrect computed patient age is showing in VX for patients less than 3 months old.
-
- Ongoing Z-1268-2024
- Ongoing Z-1269-2024
- Ongoing Z-1270-2024
- Ongoing Z-1271-2024
- Ongoing Z-1272-2024
- Ongoing Z-1273-2024
Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
-
- Ongoing Z-1171-2024
The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
-
- Terminated Z-0918-2021
The software does not update measurements and calculations in the Clinical Reporting Application (CRA) when the ventricular trace is changed in the study by a different user.
-
- Terminated Z-0282-2021
FUJIFILM has become aware of the possibility that certain CT studies may report a much higher 3D Sphere MAX value than expected. This issue is specific to 3D Sphere - other density tools are not affected.
Timeline
- Recall initiated FDA recalls Z-0918-2021
- Recall initiated FDA recalls Z-1171-2024
- Recall initiated FDA recalls Z-1268-2024
- Recall initiated FDA recalls Z-1269-2024
- Recall initiated FDA recalls Z-1270-2024
- Recall initiated FDA recalls Z-1271-2024
- Recall initiated FDA recalls Z-1272-2024
- Recall initiated FDA recalls Z-1273-2024
- Recall initiated FDA recalls Z-1407-2025
- Recall initiated FDA recalls Z-3030-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- LLZ
- Device class
- Class II
- Regulation
- 21 CFR Part 892.2050
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K190232
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| Synapse Radiology PACS Synapse Radiology PACS Software License | Synapse Radiology PACS Software License | 00854904006008 | — | On the market |
Similar devices
Other devices filed under FDA product code LLZ (System, Image Processing, Radiological).
- D2P by 3D SYSTEMS, INC.
- PRO VISION by Algotec Systems, Ltd.
- Nucleus.io by Nucleushealth, LLC
- 5D Viewer by SAMSUNG MEDISON CO., LTD.
- McKesson Cardiology by Change Healthcare Israel, Ltd.
- 21.3INCH (54CM)COLOR LCD MONITOR CCL256I2(CL21256) by Totoku Electric Co., Ltd.
- 21.3 (54 CM) MONOCHROME DIGITAL MAMMOGRAPHY LCD MONITOR, MODEL MS3112 (MDL2128A) by Totoku Electric Co., Ltd.
- hipEOS by Onefit Medical
- EndoNaut (Server) SW 1.x.x by Therenva Sas