Skip to content

Synchron Cx Systems Lipase (Lipa) Reagent

by Beckman Instruments, Inc. 510(k)

Active United States

510(k) K970905

Regulatory status

as of 28 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK970905 Date4 Jun 1997ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

1 recall · FDA

    • Terminated Z-0262-5 FDA Unclassified (pre-2012) initiated 25 Oct 2004

    Beckman Coulter has confirmed that occasionally a cuvette may be skipped during the SYNCHRON Lipase Wash (LIWA) procedure. If this occurs, there is the potential for carryover into the SYNCHRON Lipase (LIPA) assay, resulting ina substantial positive bias affecting Lipase (LIPA) results.

Specification

510(k)

510(k) Number
FDA Product Code
CHI
Device Class
Class I
Regulation
21 CFR Part 862.1465
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent

Other devices from Beckman Instruments, Inc.