Synchron Lx 20 Clinical Chemistry System, Synchron Lx Albumin Reagent, Synchron Lx Creatinine Picric Reagent, Synchron L
510(k) K965240
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK965240 | Date10 Mar 1997 | StatusActive |
Recalls
-
- Terminated Z-2130-2011
The recall was initiated because Beckman Coulter has received reports of incorrect electrolyte results on Synchron LX systems due to a variety of maintenance-related hardware issues.
-
- Terminated Z-2071-2011
This recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
-
- Terminated Z-1951-2013
Beckman Coulter is recalling the Synchron Systems LX20/UniCel DxC ISE Reference Reagent because it has a potassium (K) response near the instrument calibration flag limits and may exceed the limits on some instruments.
-
- Terminated Z-1619-2011
The recall was initiated because Cartridge Glucose (GLU) on UniCel DxC and Synchron LX instruments may give falsely high results for moderately hemolysed samples. The Impact: (1) Neonatal samples drawn by lancets are more likely to be hemolysed to a degree sufficient to cause interference with glucose estimation and are a special at-risk population, (2) Other hemolysed samples may be similar
-
- Terminated Z-2389-2010
The recall was initiated after Beckman Coulter confirmed reports that black / brown particulates were observed in some bottles of the Creatinine Alkaline Buffer (Buffer lot M911427 and M911518) in modular Creatinine (CREm) reagent kit lots M911428 and M911519. Beckman Coulter have had reports of erroneous results when using the lots of Creatinine Alkaline Buffer listed above. Single creatinine d
-
- Terminated Z-2388-2010
UniCel DxC and Synchron LX instruments produced inaccurate (low) GLUCm results. False low results could affect or delay diagnosis or treatment.
-
- Terminated Z-1094-2008
Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags or motion errors. -If a stall occurs during a GLUm test near the time of sample inject, results may be affected. -Affected GLUm results may be high or low.
-
- Terminated Z-0009-06
Synchron LX Systems Operating Software Version 4.5/v4.6 CBD (Chemistry Database) introduced an adjustment to the urine Uric Acid database calculation scheme which for some patients, depending on the volume of 24 hour urine sample, results may be reduced from above the upper limit of the laboratory's reference interval to below the limit.
-
- Terminated Z-1098-04
Numeric results are erroneously accepted by the in-house information system without qualifying symbols "<" or ">". In addition, the in-house system may interpret the results incorrectly.
-
- Terminated Z-1116-04
'Initial Rate High' parameter for User Defined Reagents does not perform as indicated in the Synchron LX Clinical Systems Operations Manual.
-
- Terminated Z-0096-04
Sample wheel may home to the wrong position.
-
- Terminated Z-0545-03
The disk that controls the home position of the sample wheel comes loose preventing a proper home sequence.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JJE
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.2160
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent