Synchron Lx Systems Immunoglobulin G/a/M Reagent

by Beckman Instruments, Inc. 510(k)

Active United States

510(k) K965108

Regulatory status

as of 29 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK965108 Date6 Mar 1997ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

2 recalls · FDA

    • Terminated Z-2226-2011 FDA Unclassified (pre-2012) initiated 11 Mar 2011 on shared 510(k) K965108

    The recall was initiated because standardization of the Synchron Systems Calibrator 1, used for both the Synchron LX and UniCel DxC systems, has been established to the new IFCC Reference Preparation for Plasma Proteins, ERM-DA470k (formerly CR470) beginning with lot number M005558. During the standardization to the new IFCC standard, ERM-DA470k, we confirmed that Complement C4 (C4) had been re

    • Terminated Z-2653-2010 FDA Unclassified (pre-2012) initiated 28 Jun 2010

    The recall was initiated after Beckman Coulter Inc. confirmed reports of high recovery of immunoglobulin A (IgA) in proficiency surveys or patient samples using Synchron Ig-A reagent lots M812164 or M902345. Using these reagent lots, some samples deficient in IgA have yielded results in the normal range. Other samples with normal levels of IgA have yielded falsely high results.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Specification

510(k)

510(k) Number
FDA Product Code
DEW
Device Class
Class II
Regulation
21 CFR Part 866.5510
Advisory Committee
Immunology
FDA Decision
SESE Substantially Equivalent