Synchron Systems Cholesterol (Chol) Reagent
by Beckman Instruments, Inc. 510(k)
Active Open recall
510(k) K971505
Regulatory status
Recall ongoing. 1 ongoing FDA recall names this device. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK971505 | Date11 Jun 1997 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-0399-2022
Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.
Specification
- 510(k) Number
- FDA Product Code
- CHH
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1175
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent