  "format": "devicerts.device.v1",
  "cite_as": "Synergy Cranial, Medtronic Navigation Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/synergy-cranial-haw (data as of 2026-11-24).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/synergy-cranial-haw",
    "json_url": "https://devicerts.com/devices/synergy-cranial-haw.json",
    "markdown_url": "https://devicerts.com/devices/synergy-cranial-haw.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-11-08",
    "status_as_of": "2026-11-24",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/synergy-cranial-haw",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Synergy Cranial",
    "registry_name": "Synergy Cranial",
    "url": "https://devicerts.com/devices/synergy-cranial-haw",
    "manufacturer": {
      "name": "Medtronic Navigation Inc.",
      "url": "https://devicerts.com/manufacturers/medtronic-navigation-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K150216",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "HAW"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K150216",
      "number_label": null,
      "device_udi_di": null,
      "date": "2015-06-11",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1666-2023",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2023-04-11",
      "reason": "In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury",
      "applies": "this device",
      "via": "K150216"
    },
    {
      "number": "Z-0383-2022",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2021-11-11",
      "reason": "Cranial biopsy procedure software can enter a state where the biopsy depth gauge is no longer synchronized with the rest of the navigational information on the screen and may display an incorrect position of the biopsy needle, which may result in prolonged procedure, the need for additional surgical procedure, tissue injury.",
      "applies": "this device",
      "via": "K150216"
    },
    {
      "number": "Z-0552-2019",
      "classification": "Class I",
      "status": "terminated",
      "initiated": "2018-09-26",
      "reason": "An incorrect display of the Biopsy Depth Gauge graphical representation in the Guidance View using Cranial software, which resulted in the biopsy of healthy tissue.",
      "applies": "this device",
      "via": "K150216"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K150216",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "HAW",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 882.4560",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Neurology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}