Syngo Imaging Xs
by Siemens AG Medical Solutions
510(k) K082430
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK082430 | Date14 Nov 2008 | StatusActive |
Recalls
-
- Terminated Z-0550-2016
For the Syngo Workflow SLR System with Software Ver: VB10C: Printouts may be printed in incorrect anatomical size when using syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an invalid combination of printer and the syngo Imaging XS printer configuration file.
-
- Terminated Z-0576-2015
Printouts may be printed in incorrect anatomical size when using syngo.plaza or syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an invalid combination of printer and the syngo.plaza printer configuration file. Printouts in anatomical size are not correct and may influence diagnostic decisions and/or therapy.
-
- Terminated Z-2111-2014
Siemens has become aware of the potential malfunction when using syngo Imaging XS, version VA70A or higher. In some cases a miscalculation may occur for the grey scale values in the functions Region of Interest, Pixel Lens, Edge enhancement and Histograms. This potential malfunction may occur for images with a pixel depth greater than 12 bit. All values calculated for images with a pixel d
-
- Terminated Z-2264-2012
There is a potential malfunction when using syngo Imaging XS, version VA70A or higher. In some situations it can happen that a miscalculation is done for the grey scale values in the functions "Regions of Interest", "Pixel Lens", "Edge Enhancement" and "Histograms".
-
- Terminated Z-1064-2012
In September 2011 firm issued two Customer Safety Advisory Notices for syngo Imaging XS. Firm became aware of an unintended behavior when using syngo Imaging XS, version VA70A and VA70B with Client/Server Installations. Loading additional images/series from the Patient Browser using the thumbnails may cause unintended deletion of images/series. In response, firm issued a Customer Safety Advisor
-
- Terminated Z-0514-2012
Unintended behavior when using syngo Imaging XS, all versions. The "Rename" functionality can cause an unintended patient merge if the renamed patient is sent to another system.
-
- Terminated Z-0515-2012
Unintended behavior when using syngo Imaging XS, versions VA70A and VA70B Client/Serve installations. Loading additional images/series from Patient Browser using the thumbnails can cause an unintended deletion of images/series.
-
- Terminated Z-1892-2010
Siemens has become aware of a potential malfunction and hazard to patients when using Syngo Imaging version lines VB20, VB30, VB35A in combination with a RIS that is violating DICOM Standard by creating non-unique study instance UIDs.
-
- Terminated Z-1423-2010
images can be overwritten
-
- Terminated Z-0174-2010
Incomplete amount of image data; specifically, The image count in the series table of the online and archive database will display the full amount of images received at the Syngo Imaging XS as archived. However, the zip file on the NAS or HSM will contain fewer images-those which have been archived until the time of the error/interruption.
-
- Terminated Z-0271-2010
- Terminated Z-0270-2010
Distorted image. The "Distorted" image cannot be used for diagnostic findings, and, depending on the customer scenario, a repeated exam could be necessary if the image cannot be reported on any other workplace or recovered from another source.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LLZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.2050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent