Syngo.via

by Siemens AG

510(k) K123375

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

8 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK123375 Date20 Nov 2012ClearedStatusActive

Recalls

8 recalls · FDA

    • Terminated Z-1778-2018 FDA Class II initiated 27 Mar 2018

    Functionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the information system. The system will still use original values.

    • Terminated Z-1261-2018 FDA Class II initiated 10 Jan 2018

    When the archiving configuration is changed, data received/created after the upgrade may be flagged as "Not to be archived". This is caused by the automatic function for cleaning up temporary data being disabled by the software upgrade. Due to the disabled cleanup function, disc capacity for free space decreases faster than usual. Unless the archiving configuration and the cleanup automatic function are reverted back to the original state (prior to the software upgrade), data that was incorrectly flagged "Not to be archived" must be manually prevented from being deleted by either clinical administrator or service engineer.

    • Terminated Z-1303-2018 FDA Class II initiated 8 Jan 2018

    A software functionality in the report sections Findings Information and Summary of Measured Findings is not performing properly. The system will use the original values instead of displaying and saving corrected/modified values. Outdated information could be sent to the referring physician when creating the formal report in the reporting system.

    • Terminated Z-0808-2018 FDA Class II initiated 31 Aug 2017

    A software update was released on August 22, 2017. Performing this software installation and upgrading syngo.via system to the software version VB20A_HF02 will resolve various bugs and inconsistent diagnosis with the old system.

    • Terminated Z-1689-2017 FDA Class II initiated 6 Mar 2017

    Software changes now available to address several issues

    • Terminated Z-2853-2016 FDA Class II initiated 5 Aug 2016

    Software error. Incorrect values for the volume calculation from a freehand VOI at the customer site. In volume calculations of prostate as well as in volume calculations of liver were too high.

    • Terminated Z-2245-2016 FDA Class II initiated 23 Jun 2016

    Incorrect values for the volume calculation. Software update VB30B via Update Instructions SY018/16/P to resolve software errors.

    • Terminated Z-1705-2015 FDA Class II initiated 15 May 2015

    measurements drawn on the 2nd and subsequent images of the series are not visible on printouts when the series is sent to print.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LLZ
Device Class
Class II
Regulation
21 CFR Part 892.2050
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent