Synthes Locking Distal Radius Plating System
by Synthes (Usa) 510(k)
Active
510(k) K012114
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK012114 | Date28 Sep 2001 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2389-2016
It was reported that the specified lots of the 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm can break between the head and the shaft. This resulted from the neck feature being too close to the StarDrive feature leaving a thin wall between neck and drive.
-
- Terminated Z-0323-2016
It was discovered that the above part number and lot of 2.4mm Locking Screws may contain the incorrect package insert (GP2613, VA-LCP Distal Radius System). The correct package insert for the above part is GP2615 (Modular Mini Fragment LCP System).
-
- Terminated Z-0489-2007
Hole is malformed, leading to reduced construct strength.
Specification
- 510(k) Number
- FDA Product Code
- HRS
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3030
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent