Synthes Mirs
by Synthes Spine 510(k)
Active
510(k) K113044
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK113044 | Date21 Dec 2011 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0396-2014
- Terminated Z-0395-2014
This action is being initiated following a detailed investigation in response to reported complaints of post-operative locking cap loosening. Synthes has concluded that if the construct is mis-aligned during final tightening, the potential exists for locking caps to loosen post-operatively in certain situations.
-
- Terminated Z-0001-2013
Synthes is initiating a Medical Device Labeling Correction following a detailed review of the current labeling in response to reported complaints. Synthes has identified technique measures and has updated the labeling associated with this system. The labeling reflects updated technique recommendations and emphasizes the importance of avoiding rod mis-alignment.
Specification
- 510(k) Number
- FDA Product Code
- NKB
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3070
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent