Synthes Sternal Zipfix System
510(k) K110789
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK110789 | Date28 Jul 2011 | StatusActive |
Recalls
-
- Terminated Z-0276-2016
Certain lots of the Application Instrument for Sternal ZIPFIX that were previously recalled on August 19, 2015 and subsequently repaired and returned to customers between August 26, 2015 and September 23, 2015. The screws on the repaired instruments have the potential to break or loosen. DePuy Synthes is requesting the return of the repaired device in exchange for a new device.
-
- Terminated Z-2475-2015
The end cap may loosen and detach making the instrument non-functional. No injuries reported.
-
- Terminated Z-0114-2014
When using the 1st generation Zip Fix Application Instrument it is possible to cut the implant while the tensioning trigger is being squeezed which could lead to loosening of the implant.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JDQ
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3010
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent