Table, X-Ray, Rfx Classical
510(k) K760466
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK760466 | Date26 Aug 1976 | StatusActive |
Recalls
-
- Terminated Z-2179-2017
Reported incidents of a patient step detaching from the table. A fall from a patient step detaching while in use could result in an injury to a patient or operator.
-
- Terminated Z-0915-2017
GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors - Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems.
-
- Terminated Z-0012-2014
GE Healthcare has recently become aware of a potential safety issue involving the spotfilm device power assist handle. There is a potential for the internal bolts on the power assist handle to be loose or to loosen and/or fall out. This issue, if present, can result in a loss of calibration and offset of the power assist handle and/or, under extreme condition, sticking of the handle in the ON p
-
- Terminated Z-0200-2009
GE Healthcare has recently become aware of a potential risk of monitor dropping associated with the monitor suspension of the Advantx system that may impact patient safety. It has been reported that suspensions manufactured after March 8, 1995 did not have the required thread locking agent applied to the setscrew, allowing the setscrew to back out over time, which can lead to the monitor dropping
-
- Terminated Z-0853-04
- Terminated Z-0852-04
- Terminated Z-0851-04
- Terminated Z-0850-04
The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IXR
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1980
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent