Tandem Mobi Mobile Application, iOS 1016909
by Tandem Diabetes Care, Inc. EUDAMEDGUDID
Active
Ref 1016909
Regulatory status
Recall ongoing. 3 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | NumberMDR 824766 | Date2 Sep 2026 | StatusActive |
| United States | GUDID listing GUDID | Number00389152690940 | Date6 Oct 2025 | StatusIn distribution |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
- European Union
- United States
Certificates
- MDR 824766 Supplemented
Technical documentation EU MDR
- Notified body
- Notified body 2797
- Issued
- 2 Sep 2026
- Expires
- 14 May 2030
- Reason
- Change approved
Recalls
-
- Ongoing Z-0427-2026
Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.
-
- Ongoing Z-0232-2026
A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.
-
- Ongoing Z-1520-2025
A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.
Specification
- Trade Name
- Tandem Mobi Mobile App EUDAMEDTandem Mobi Mobile Application, iOS GUDID
- Generic Name
- Ambulatory insulin infusion pump, electronic, software-dosing, scalar algorithm
- Device Name
- Tandem Mobile Application
- Model / Reference
- 1016909
- Primary DI / UDI-DI
- 00389152690940
- Basic UDI-DI
- 0850018992TF-0011603KR
- Direct Marketing DI
- 00389152690940
- 510(k) Number
- FDA Product Code
- QFG
- EMDN Code
- Z12030382 Infusion instruments - software accessories
- GMDN Term
- Ambulatory insulin infusion pump, electronic, software-dosing, scalar algorithm
- Device Class
- Class III EUDAMEDClass II GUDID
- Regulation Number
- 880.5730
- Special Device Type
- Software
- Market of Registration
- Belgium
- Commercial Distribution
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device Class | Class III | Class II |
Other devices from Tandem Diabetes Care, Inc.
Properties
Other properties (21)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No