# TCM400

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/tcm400-lpp>
- Manufacturer: [Radiometer Medical A/S](https://devicerts.com/manufacturers/radiometer-medical-aps) (FDA registration: Denmark)
- Device class: Class II (US FDA)
- FDA product code: LPP

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K001866 · decision 2000-12-07 · decision SESE · status: Active

## Recalls

- Z-1755-2020 · Class II · terminated · initiated 2020-02-20 · matched through shared submission K001866, may concern other models: Reports have been received about the transcutaneous monitoring system display flickering or monitor becoming non responsive and beeping. This may result in delayed patient monitoring as well as potentially compromise the product performance.
- Z-1930-2013 · Class II · terminated · initiated 2012-11-27: RADIOMETER became aware that the TCM4 series Base Unit does not comply with IEC 60601-1 Sub Clause 15B.
- Z-1436-06 ·  · terminated · initiated 2005-11-01 · matched through shared submission K001866, may concern other models: Excessive Drift-the membrane units of the device cause the electrode to exceed performance standards for drift in the first two or more calibrations

## Identifiers

- 510(k) Number: K001866 [FDA 510(k)]
- FDA product code: LPP [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 868.2500 [FDA 510(k)]
- Advisory committee: Anesthesiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: TCM400, Radiometer Medical A/S. deviCERTS medical device catalogue. https://devicerts.com/devices/tcm400-lpp (data as of 2026-11-20).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-20
- JSON: <https://devicerts.com/devices/tcm400-lpp.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/tcm400-lpp
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.
