Tdx Theophylline
by Abbott Laboratories 510(k)
Active
510(k) K833070
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK833070 | Date14 Nov 1983 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0958-2007
A decrease in mP values (calibration curve Net Polarization value) has been observed with this reagent lot at all calibrator levels and may produce the following results: Calibration Error Code 148 (Polarization Too Small), Calibration Error Code 172 (Calibration Not Stored or Calibration Not Accepted) Controls out of range (when using a stored calibration curve).
Specification
- 510(k) Number
- FDA Product Code
- LER
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.3880
- Advisory Committee
- Clinical Toxicology
- FDA Decision