Temperature Control & Monitor
by Sams, Inc.
510(k) K841402
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK841402 | Date11 Jun 1984 | StatusActive |
Recalls
-
- Terminated Z-1790-2021
Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.
-
- Terminated Z-1417-2018
Update to cooler-heater cleaning instructions.
-
- Terminated Z-2208-2016
Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DWC
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.4250
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent