Tempus LS - Manual
by Remote Diagnostic Technologies, Ltd. A Philips Company
510(k) K200849
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK200849 | Date24 Jul 2020 | StatusActive |
Recalls
-
- Ongoing Z-0557-2025
Due to incorrect labels applied to devices during servicing. Labels contain the incorrect product name and CAT number.
-
- Ongoing Z-2028-2024
Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or ineffective pacing, or no or inappropriate defibrillation pulse.
-
- Ongoing Z-1825-2023
Defibrillator/pacemaker module (DPM) may encounter a problem causing communication from the Mainboard to the DPM to fail, leading to no or ineffective pacing with displayed error message "HW failure DPM".
-
- Terminated Z-0564-2021
A software error was detected within software version 1.3.4 for the Tempus LS device. When exiting pacer back into manual mode the pacer pulse wave may put the ECG calibration out of calibration. This software error may result in a delay in therapy, as the clinician is not provided an ECG waveform on which a shockable rhythm decision can be made. The device is still able to deliver a manual defibrillation shock.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LDD
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.5300
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent