# Tf TF7197

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/tf-tf7197>
- Manufacturer: [Randox Laboratories, Ltd.](https://devicerts.com/manufacturers/randox-laboratories-ltd)
- Name in the registry: TF TF7197
- Device class: Class II (US FDA)
- FDA product code: DDG
- GMDN: Transferrin/transferrin isoform IVD, kit, nephelometry/turbidimetry (59041)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 05055273206500 · listed 2016-09-24 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K002585 · decision 2001-02-05 · decision SESE · status: Active · from GUDID

## Recalls

- Z-1422-2022 · Class III · terminated · initiated 2022-06-09: A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received.
- Z-1423-2022 · Class III · terminated · initiated 2022-06-09: A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received.

## Identity

- Trade name: TF [FDA GUDID (AccessGUDID)]
- Generic name: Transferrin/transferrin isoform IVD, kit, nephelometry/turbidimetry [FDA GUDID (AccessGUDID)]
- Model / Ref / Catalog: TF7197 [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 05055273206500 [FDA GUDID (AccessGUDID)]
- 510(k) Number: K002585 [FDA GUDID (AccessGUDID)]
- FDA product code: DDG [FDA GUDID (AccessGUDID)]
- GMDN term: Transferrin/transferrin isoform IVD, kit, nephelometry/turbidimetry (59041) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 866.5880 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Single use: Yes [FDA GUDID (AccessGUDID)]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Convenience kit: Yes [FDA GUDID (AccessGUDID)]
- Lot / batch number: Yes [FDA GUDID (AccessGUDID)]
- Expiration date: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Serial number: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Tf TF7197, Randox Laboratories, Ltd. deviCERTS medical device catalogue. https://devicerts.com/devices/tf-tf7197 (data as of 2026-11-25).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-25
- JSON: <https://devicerts.com/devices/tf-tf7197.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/tf-tf7197
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.