The NOxBOXi Nitric Oxide Delivery System
510(k) K201339
Active
Valid until
No expiry
FDA recalls
4 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK201339 | Date19 Jun 2020 | StatusActive |
Recalls
-
- Ongoing Z-3306-2026
Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors.
-
- Ongoing Z-3165-2026
Potential for oxygen tubing leaks due to failure of the tubing or tubing connection, which may lead to oxygen desaturation.
-
- Ongoing Z-0028-2026
Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
-
- Ongoing Z-0029-2026
The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MRN
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.5165
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent