Skip to content

Tinaquant Complement C4 ver2 5991994190

by Roche Diagnostics GmbH

Tinaquant Complement C4 ver2 5991994190 is a Class II device made by Roche Diagnostics GmbH, cleared under FDA 510(k) K012359 on 29 Nov 2001 and listed in FDA GUDID since 20 Sep 2016 (UDI-DI 04015630928460), active as of 12 Nov 2026. One FDA recall is linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

1 closed

Documents

510(k) K012359

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 12 Nov 2026.

Regulatory status

as of registry sync 12 Nov 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number04015630928460 UDI-DI Date20 Sep 2016ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK012359 Date29 Nov 2001DecidedStatusActive

Official records

Across registries

2 registries list this device

  • FDA 510(k) K012359 decided 29 Nov 2001 Active
  • FDA GUDID UDI-DI 04015630928460 listed 20 Sep 2016 Listed

Recalls

1 recall · FDA

    • Terminated Z-0103-2021 FDA Class III initiated 27 Aug 2020

    Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.

Timeline

3 dated events

  1. Cleared FDA 510(k) K012359
  2. Listed FDA GUDID 04015630928460
  3. Recall initiated FDA recalls Z-0103-2021

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
Tinaquant Complement C4 ver2
Generic name
Complement components C4/C4c IVD, kit, nephelometry/turbidimetry
Model / Ref / Catalog
5991994190
Primary DI / UDI-DI
04015630928460
510(k) Number
FDA product code
DBI
GMDN term
Complement components C4/C4c IVD, kit, nephelometry/turbidimetry
Device class
Class II
Regulation number
866.5240
GUDID distribution status
In Commercial Distribution

Other devices from Roche Diagnostics GmbH

See all devices from Roche Diagnostics GmbH →

Properties

  • Single use
  • Prescription use
  • Lot / batch number
  • Expiration date
Other properties (11, all answered No)