Tissue Approximation System (TAS)
by TAS Medical, Inc. 510(k)
Active
510(k) K220980
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK220980 | Date5 Jul 2023 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-1294-2025
Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.
Specification
- 510(k) Number
- FDA Product Code
- GAR
- Device Class
- Class II
- Regulation
- 21 CFR Part 878.5020
- Advisory Committee
- General, Plastic Surgery
- FDA Decision
- SESE Substantially Equivalent