ToggleLoc 904722
ToggleLoc 904722 is a Class II device made by Biomet Orthopedics, Inc., cleared under FDA 510(k) K130033 on 1 Aug 2013 (3 earlier submissions) and listed in FDA GUDID since 24 Oct 2015 (UDI-DI 00880304453289), active as of 23 Oct 2026. 9 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00880304453289 | Date24 Oct 2015 | StatusListed |
| United States | FDA 510(k)SESE | NumberK130033 | Date1 Aug 2013 | StatusActive |
Official records
- AccessGUDID record 00880304453289
- FDA 510(k) database K130033
- FDA 510(k) database K083070
- FDA 510(k) database K033838
- FDA 510(k) database K973775
Across registries
- FDA 510(k) K130033 decided 1 Aug 2013 Active
- FDA GUDID UDI-DI 00880304453289 listed 24 Oct 2015 Listed
Recalls
-
- Ongoing Z-0212-2026
There is a potential for a missing slotted button assembly in one lot of ZipTight, Acute AC Joint Implant, Single Ziploop
-
- Terminated Z-0818-2018
Ziptight Ankle Syndesmosis Fixation was packaged without the tip protectors.
-
- Terminated Z-1828-2016
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.
-
- Terminated Z-1363-2015
The firm is recalling ToggleLocs and Ziploops manufactured in May and July of 2014 as they were not processed following a standard work process, which resulted in complaints of sterile packaging not being sealed.
-
- Terminated Z-1953-2010
- Terminated Z-1952-2010
The firm has received complaints that the saddle is too long. The saddle is supposed to be between 7/8 and 1 1/8 inches long. The saddle in the complaints was up to 1.9 inches long.
-
- Terminated Z-0560-2010
- Terminated Z-0559-2010
Packages labeled as containing a stainless steel device actually contain a titanium device, and vice versa.
-
- Terminated Z-0582-2012
Six sizes of ToggleLoc, Artificial Ligament Fixation Device, have been manufactured with the suture pull hole drilled in the wrong location on the device. The ToggleLoc implant is deployed with a suture that pulls the device through the tunnel. The hole being drilled in the wrong location will orient the device to pass through the tunnel with the angled end of the implant first. This may caus
Timeline
- Recall initiated FDA recalls Z-0560-2010
- Recall initiated FDA recalls Z-0559-2010
- Recall initiated FDA recalls Z-1953-2010
- Recall initiated FDA recalls Z-1952-2010
- Cleared FDA 510(k) K130033
- Recall initiated FDA recalls Z-1363-2015
- Listed FDA GUDID 00880304453289
- Recall initiated FDA recalls Z-1828-2016
- Recall initiated FDA recalls Z-0818-2018
- Recall initiated FDA recalls Z-0212-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- ToggleLoc
- Generic name
- Tendon/ligament bone anchor, non-bioabsorbable
- Model / Ref / Catalog
- 904722
- Primary DI / UDI-DI
- 00880304453289
- FDA product code
- JDR, MBI
- GMDN term
- Tendon/ligament bone anchor, non-bioabsorbable
- Device class
- Class II
- Regulation number
- 888.3030, 888.3040
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Conditional
Similar devices in this category
Other devices filed under Tendon/ligament bone anchor, non-bioabsorbable.
- Titanium Interference Screw 10011 by PARCUS MEDICAL LLC
- Endochor PLT-EN4HP1625 by Bonegraft Biomaterials Co.
- Bonetech C-SCRW-BOBS15310 by Bonegraft Biomaterials Co.
- PEEK CF Interference Screw 10027 by PARCUS MEDICAL LLC
- TENODESIS SCREW 7X20 MM TM50730720 by Tulpar Tıbbi Ürünler İthalat İhracat Sanayi ve Ticaret L.T.D Ş.T.İ
- TENODESIS SCREW 10X27 MM TM00733027 by Tulpar Tıbbi Ürünler İthalat İhracat Sanayi ve Ticaret L.T.D Ş.T.İ
- PULLUP CLIP PULL0CT201 by S.B.M
- 10011 by Innovate Orthopaedics, Ltd.
- Inion Hexalon™ Biodegradable ACL/PCL Screw SET-4000 by Inion, Ltd.
- Inion Hexalon™ Biodegradable ACL/PCL Screw ACL-4109 by Inion, Ltd.