Total Iron Binding Capacity (Ibct) Flex Reagent Cartridge
by Dade Behring, Inc. 510(k)
Active
510(k) K994115
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK994115 | Date28 Jan 2000 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0933-2013
The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid. Siemens is conducting a voluntary corrective action for Dimension Total Iron Binding Capacity (IBCT) to modify the analytical sensitivity claim to 36 ug/dL (6.44 umol/L).
-
- Terminated Z-2385-2010
Test produces falsely elevated IBCT results and Abnormal Reaction test report messages on heparinized plasma samples.
Specification
- 510(k) Number
- FDA Product Code
- JMO
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1415
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent