TrackX VISION-T SNAP
TrackX VISION-T SNAP (TX-3-3108) is a medical device (FDA Class II; FDA product code OWB, Interventional Fluoroscopic X-Ray System) made by TRACKX TECHNOLOGY Inc., cleared under FDA 510(k) K173736 on 28 Feb 2018 and listed in FDA GUDID since 9 Oct 2018 (UDI-DI 00857088008042), active as of 22 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00857088008042 | Date9 Oct 2018 | StatusListed |
| United States | FDA 510(k)SESE | NumberK173736 | Date28 Feb 2018 | StatusActive |
Official records
- AccessGUDID record 00857088008042
- FDA 510(k) database K173736
From the FDA record
- Device description
- Disposable snap that snaps onto a Jamshidi for the purpose of instrument tracking. Component body part # TX-3-3203
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K173736 decided 28 Feb 2018 Active
- FDA GUDID UDI-DI 00857088008042 listed 9 Oct 2018 Listed
Recalls
-
- Ongoing Z-0260-2025
The patient label stickers may have the wrong expiration date recorded. The contract manufacturer entered the manufacturing date (2024-04-01) as the expiration date. The actual expiration date of the product is 2027-04-01. This is reflected on the carton label and sterile pouch label correctly.
-
- Terminated Z-1107-2023
The device is breaking and falling off the Nuvasive T-Handle with Hudson Adapter on certain Nuvasive XLIF instruments during surgery when malleted.
Timeline
- Cleared FDA 510(k) K173736
- Listed FDA GUDID 00857088008042
- Recall initiated FDA recalls Z-1107-2023
- Recall initiated FDA recalls Z-0260-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- TrackX
- Generic name
- Diagnostic x-ray digital imaging system workstation application software
- Model / Ref / Catalog
- VISION-T SNAP, TX-3-3108
- Description
- Disposable snap that snaps onto a Jamshidi for the purpose of instrument tracking. Component body part # TX-3-3203
- Primary DI / UDI-DI
- 00857088008042
- 510(k) Number
- FDA product code
- OWB
- GMDN term
- Diagnostic x-ray digital imaging system workstation application software
- Device class
- Class II
- Regulation number
- 892.1650
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code OWB (Interventional Fluoroscopic X-Ray System).
- PHILIPS DUO DIAGNOST by Philips Medical Systems North America, Inc.
- SYNC-RX SYSTEM by VOLCANO CORPORATION
- GE INNOVA SOLID STATE XRAY-IMAGER FLUOROSCOPIC X-RAY SYSTEM, MODEL: 4100IQ, 3100IQ, 2100IQ by GE Healthcare
- EmboGuide by Philips Medical Systems Nederland B.V.
- OEC 9900 ELITE by Ge Healthcare Surgery
- DISCOVERY IGS 740 by GE Medical Systems SCS
- ANT-X System by Ndr Medical Technology Pte., Ltd.
- OEC Elite by GE OEC MEDICAL SYSTEMS, INC
- Allia IGS 3, Allia IGS 5, Allia IGS 7, Allia IGS 7 OR by GE Medical Systems SCS
- Cios Select (VA21) Flat Panel by SIEMENS MEDICAL SOLUTIONS USA, INC.